Recall Z-1119-2015 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis One; The Artis One is an angiography system developed for diagnostic interventional procedures including, but not limited to pediatric and obese patients. Procedures that can be performed with the Artis One include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as proced

DateFeb 17, 2015 (initiated Jan 16, 2015)
StatusTerminated
ReasonThe possibility exists that the monitor may fail and requires a power circle (shutdown and then power on) to resume operation. The problem is not systematic; but sporadic on single units.
Root causeProcess control
ActionSiemens notified their customer via Customer Safety Advisory Letter AX049/14/S, dated January 16, 2015, and has initiated a fix with Update Instruction AX048/14/S which will be completed by a Siemens service technician. Customers with questions should contact the Customer Care Center at 1-800-888-7436. For questions regarding this recall call 610-219-6300.
Quantity1
DistributionNationwide Distribution to MI only
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133580ARTIS ONE
Siemens Medical Solutions USA, Inc., Apr 28, 2014