Recall Z-1283-2015 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Stryker Howmedica Osteonics Corp.

Triathlon Distal Capture Assembly Catalog No: 6541-1-723 The Triathlon Distal Capture Assembly is used when a surgeon elects to use a capture for the distal femoral resection or proximal tibial resection in a Triathlon primary total knee arthroplasty (TKA).

DateMar 16, 2015 (initiated Feb 12, 2015)
StatusTerminated
ReasonPotential for weld disassociation of the cross pin, which could lead to loose or disassociated trigger mechanism on the Distal Capture Assembly.
Root causeDevice Design
ActionOn March 20, 2015, Stryker sent Urgent Medical Device Recall Notifications/Urgent Medical Device Recall Notification Acknowledgement Forms dated March 19, 2015 to all affected customers. The letter identified the product, the related issue and the action to be taken by the customers. Customers were asked to return the attached acknowledgement form confirmed they received and reviewed the notification. Customers with questions were instructed to call 201-831-6365. Stryker sent an Urgent Medical Device Correction letter dated February 11, 2015, to all affected customers. The letter identifi
Quantity2030 units
DistributionWorldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Canada, Hong Kong, Netherlands, Sweden, Switzerland, Spain, France, Italy, UK, India, Australia, Korea, Panama, Polan, Ireland, Kuala Lumpar, New Zealand and South Africa.