ClearanceBrief
Recall Z-1066-2015 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment

Stryker Instruments Div. of Stryker Corporation

System 6 Aseptic Housing. The Stryker System 6 Aseptic Housings provide a sterile enclosure for the Stryker Non-sterile Battery. Accessory for powered orthopedic surgical instrument.

DateFeb 9, 2015 (initiated Jan 7, 2015)
StatusTerminated
ReasonStryker is recalling System 6 Aseptic Housing, due to a failed Continuous Bond Line Test from a housing without a continual weld. Customers are instructed to return recalled product to Stryker.
Root causeUnder Investigation by firm
ActionStryker sent notice to customers via Fed ex overnight on 1/16/2015, regarding its voluntary recall of Stryker System 6 Aseptic Housings because the laser welder may have failed to produce a continuous bond line for the four specific lots: 13205, 13209, 13210, 13212. Customers are instructed to discontinue use, complete the business reply form, and return all recalled product. Customers may contact Kara Spath at 269-389-4518, or kara.spath@stryker.com.
Quantity399
DistributionWorldwide Distribution-US (nationwide) including Puerto Rico and the states of AR, CA, CO, FL, GA, HI, MD, NC, NY, PA, SC, TX, and VA, and the countries of Australia, Canada, Switzerland, Hong Kong, South Africa, Chile, Korea, Poland, India, Colombia, The Netherlands, and France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.