Recall Z-1076-2015 ยท FTL Mesh, Surgical, Polymeric

MAST Biosurgery USA, Inc

The SurgiWrap Bioresorbable Adhesion Barrier Film. SurgiWrap 100x130x0.05mm (01) 18964050255027 (17) 170731(10)52804 LOT 52804 2017-08 REF25502-01 Product is wrapped in a foil pouch that is then sealed into an additional envelope.

DateFeb 10, 2015 (initiated Aug 26, 2014)
StatusTerminated
ReasonThe expiration date of the bar code on the envelope label did not match the expiration date next to the hourglass. The printed expiration date was August of 2017, while the bar code date read as July 31, 2017. The correct expiration date was July 31, 2017.
Root causeLabeling Change Control
ActionMAST Biosurgery labeled 49 new envelopes and sent them to the distributor, and subsequently had the distributor remove the contents of the outer envelope (instructions for use and the foil pouches containing the product) and place them into the newly supplied and properly labeled outer envelopes provided by them. The distributor requested to send back the original 49 enveloped. The second batch of new labels was inspected and attached to the new outer envelope to create a properly labeled outer envelope prior to shipping. The firm requested that once the distributor replaced the product an
Quantity49 pieces
DistributionWorldwide Distribution to the country of China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061473SURGI-WRAP MAST BIORESORBABLE SHEET