Recall Z-1215-2015 ยท OLO Orthopedic Stereotaxic Instrument

Brainlab AG

The Brainlab Offset Cup Impactor Universal. Part of the Brainlab hip system. An intraoperative image-guided localization system to enable minimally invasive orthopedic surgery. (a stereotaxic instrument)

DateFeb 26, 2015 (initiated Jan 16, 2015)
StatusTerminated
ReasonThe recommended sterilization and drying parameters are not effective to achieve the minimum required sterilization level.
Root causeReprocessing Controls
ActionFIELD SAFETY NOTICE / PRODUCT NOTIFICATION letters dated December 11, 2014 were sent to all direct accounts (customers). The letters included instructions for customers to utilize the sterilization cycles included in the letter. Customers were further advised to alert users of these devices to the sterilization process changes. Customers with questions can contact Brainlab by telephone at 800-597-5911 or via e-mail at us.support@brainlab.com.
Quantity114 offset cup impactors
DistributionWorldwide Distribution - US: Nationwide (AZ, CA, ID, IL, MD, MI, MN, OH, OK, OR, TN, WA, WI, WV); Australia, Austria, Bahrain, Belgium, Germany, Hong Kong, India, Italy, Japan, Malaysia, Russian Federation, Saudi Arabia, South Korea, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122011NAVIGATION SOFTWARE HIP-UNIVERSAL; NAVIGATION SOFTWARE HIP THR-AESCULAP
Brainlab AG, Nov 15, 2012