ClearanceBrief
Recall Z-1278-2015 ยท KXG Applicator, Absorbent Tipped, Sterile

Puritan Medical Products Co., LLC

Puritan absorbent foam tipped applicator; Catalog Number 25-1406 1 PF BT. Sterile absorbent foam tipped applicator used to apply medication or to obtain specimens from a patient.

DateMar 13, 2015 (initiated Dec 22, 2014)
StatusTerminated
ReasonShaft breaks so that adequate patient specimen sample cannot be obtained.
Root causeProcess control
ActionConsignees were initially notified of the recall via phone on 12/23/2014 and 12/29/2014. They were informed that the product may have been over exposed to gamma radiation, resulting in brittle sticks. They were instructed to not use the product for patient sampling and to return any product remaining in inventory to Puritan.
Quantity9 cases of 500 applicators each
DistributionDistributed in the states of MS, ME, GA, and NC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall