Recall Z-1079-2015 ยท KXJ Table, Radiologic

Villa Radiology Systems LLC

Juno DFR X-ray system

DateMar 6, 2015 (initiated Dec 14, 2014)
StatusTerminated
ReasonThe system does not provide the appropriate audible signal, permanent activation, and manual override, although the system is in high-level control functionality
Root causeRadiation Control for Health and Safety Act
ActionPlanned Action by the firm: 1. You will send out a customer notification letter that identified the software corrective actions and that indicates that these corrective actions will be provided to the end-user free of charge. 2. You will implement a corrective software patch so that kV and mA manual adjustments from the generator console are inhibited, and consequently no high-level control modality is allowed. In AERC, high-level control is already disabled. A Field Service Engineer is to implement the software patch according to the corrective action plan 1. Notification of all dealers a
Quantity30 units distributed in the US
DistributionUS Distribution to the states of: AZ, AL, CA, CO, CT, DC, FL, IA, IL, IN, MI, MD, MS, MA, MO, NY, OH, PA, UT, TN and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050190APOLLO
Villa Sistemi Medicali S.P.A., Mar 25, 2005