Recall Z-1061-2015 ยท LNH System, Nuclear Magnetic Resonance Imaging

Alltech Medical Systems America Inc

Echostar Spica 1.5T MRI system, for magnetic resonance imaging of the body.

DateFeb 18, 2015 (initiated Jan 9, 2015)
StatusTerminated
ReasonThe MRI system safety feature, which lowers the magnetic field in emergency situations, may not work correctly.
Root causeProcess control
ActionAltech Medical Systems America Inc., notified customers on January 9, 2015, of the problem via telephone. Customers were instructed to immediately cease using the product until the correction can be made. A follow-up letter was sent through the mail, which stated that a service engineer would make a site visit to ramp down the magnet until the issue can be corrected. Customers were asked to complete the attached Acknowledgement and Receipt Form and fax it to (440) 424-2255. For questions contact AMSA at 1-855-200-2672.
Quantity2 units
DistributionUS Distribution in the state of Maryland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113511ECHOSTAR SPICA 1.5T MR SYSTEM
Alltech Medical Systems America, Inc., May 2, 2012