ClearanceBrief
Recall Z-1040-2015 ยท IZL System, X-Ray, Mobile

Sedecal S.A.

MobileDiagnost wDR motorized portable diagnostic X-ray systems.

DateFeb 2, 2015 (initiated Aug 6, 2014)
StatusTerminated
ReasonReports of unexpected movement.
Root causeOther
ActionURGENT - Field Safety Notice letters dated August 6, 2014 were sent to all consignees of record. The letters instructed consignees: 1) The unit is equipped with a safety switch in the drive handle. If the drive handle is released the unit will stop; 2) Do not try to correct unintended movement using the controls on the drive handle; 3) Instead, stop the system immediately by releasing the drive handle entirely; 4) Alternatively, the emergency button can be pressed; 5) Take special care when the system crosses metal strips or plates on the floor or enters an elevator, as these are situations wh
Quantity283 systems
DistributionUS: Nationwide (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111725MOBILEDIAGNOST WDR
Sedecal., Sa., Jul 19, 2011