Recall Z-1031-2015 ยท FKX System, Peritoneal, Automatic Delivery
Baxter Healthcare Corp.
HomeChoice/HomeChoice Pro Automated Personal Cycler, for Automated Peritoneal Dialysis. Product codes 5C8310, 5C8310R, R5C8320, and 5C4474.
| Date | Jan 29, 2015 (initiated Oct 21, 2014) |
| Status | Terminated |
| Reason | The keypad buttons on HomeChoice devices may be activated without the operator pressing them. |
| Root cause | Nonconforming Material/Component |
| Action | An Urgent Device Correction notification was sent to consignees on October 31, 2014. Within the notification, Baxter Healthcare asked customers to do the following: 1) Contact Baxter Technical Services to arrange for the exchange (swap) of any affected HomeChoice PRO Automated PD System devices. 2) Complete the attached customer reply form and return it to Baxter by either fax or scanned e-mail. For general questions regarding this communication, contact The Center for One Baxter at 900-422-9837, Monday through Friday between the hours of 8:00 AM and 5:00 PM Central Time. |
| Quantity | 200 |
| Distribution | AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI and District of Columbia |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K102936 | HOMECHOICE/HOMECHOICE PRO AUTOMATED PERSONAL CYCLER PERITONEAL DIALYSIS SYSTEM MODEL 5C4471, 5C8310, 5C4471R AND 5C8310R Baxter Healthcare Corp, Mar 30, 2011 |