Recall Z-1035-2015 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare

This correction applies to all GE Healthcare MRI systems with superconducting magnets: Discovery MR450, Discovery MR750, Signa HDx, Signa HDxt, Signa HDi, GE 1.5T Signa HDe, Optima MR360, Brivo MR355, Signa Excite-HD 1.5T & 3T, Signa Excite 1.5T HD Twinspeed, Signa Excite 1.5T HD Echospeed, Signa Excite 1.5T HD Highspeed, 1.5T Signa Infinity TwinSpeed , 1.5T Signa Infinity EchoSpeed Plus, 1.5T Sig

DateFeb 18, 2015 (initiated Jan 7, 2015)
StatusTerminated
ReasonAt certain sites, the MRU may not be connected to the magnet. In emergency situations, a disconnected MRU could delay removal of a ferrous object from the magnet, potentially resulting in life-threatening injuries. The MRU must be connected to the magnet at all times.
Root causeEmployee error
ActionGE Healthcare sent an "Urgent Medical Device Correction" letter GEHC Ref# 60876 dated January 6, 2015 to affected consignees. The letter was addressed to Hospital Administrators / Risk Managers, Radiology Department Managers, & Radiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction & Contact Information. Customers were instructed to do the following: As a preventative measure, confirm that MRU is connected to the magnet by performing the following four step test on the MRU. 1. Verify the green CHARGER POWER LED is lit. 2. Depre
Quantity12,968 (5,708 US, 7,260 OUS).
DistributionWorldwide Distribution - US Nationwide in all states in continental USA including DC, PR, GU, and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BOLIVIA, BOSNIA & HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0122001.5T INFINITY TWINSPEED MR SYSTEM
GE Medical Systems, Aug 3, 2001
K013636SIGNA INFINITY MR SYSTEM WITH EXCITE TECHNOLOGY
GE Medical Systems, Jan 24, 2002
K030874GE 3.0T SIGNA INFINITY TWINSPEED WITH EXCITE MR SYSTEM
General Electric Co., Mar 28, 2003
K040444GE SIGNA 3.0T WITH EXCITE MR SYSTEM
General Electric Co., Apr 14, 2004
K041476GE SIGNA EXCITE 1.5T MR SYSTEM, GE SIGNA EXCITE 3.0T MR SYSTEM
Ge Medical Systems, LLC, Jun 17, 2004