Recall Z-2069-2015 ยท CAN Regulator, Pressure, Gas Cylinder

Western / Scott Fetzer Company

Various OxyTOTE Portable Oxygen System regulators, Valve Integrated Pressure Regulator, one unit per package. Product Usage: Used to dispense Oxygen at prescribed flow rates to patients for therapeutic purposes and is designed to provide mobility with a self-contained supply of Oxygen.

DateJul 20, 2015 (initiated Jan 31, 2015)
StatusTerminated
ReasonThe firm has become aware of the possibility of ignition and rupture (explosion) of a compressed gas oxygen cylinder, which can cause serious injury and/or death.
Root causeComponent design/selection
ActionWestern/Scott Fetzer sent an Urgent Medical Device Recall Notice notification dated January 31, 2015 to their customers via U.S. mail. The notification identified the affected product, problem and actions to be taken. The OxyTOTE/oxyQuik/AirTOTE Distributors and Oxygen Service Providers were instructed to: 1) Notify and provide the letter to all accounts/customers to which any OxyTOTE/oxyQuik/AirTOTE products have been distributed. Alternately, please provide a list of customers who received affected product and Western/Scott Fetzer will notify them on your behalf. 2) Referring to the in
Quantity162,678 units
DistributionWorldwide Distribution -- US, including the states of Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Ca
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.