ClearanceBrief
Recall Z-1306-2015 ยท MMP Cover, Barrier, Protective

Penner Mfg Inc

Penner Patient Stretcher/Transfer Lift System Models 391000-1 and 392000-1 The Pacific Stretcher/Transfer Lift is used with Whirlpool or Aqua-Aire (air bubbling) bathing system intended for use in nursing homes, hospitals, and assisted living facilities to transfer and/or lift patients under the direct supervision of trained staff. Model 392000-1 is equipped with a scale.

DateMar 20, 2015 (initiated Jan 30, 2015)
StatusTerminated
ReasonIf the patient sits on either wing, not squarely on the center section of the Stretcher Lift, tipping and injury to the operator or patient could result. The device manual was updated with an additional warning statement.
Root causeUse error
ActionBeginning on 1/30/2015, the recalling firm sent letters via certified mail to their customers informing them of the change to the stretcher manual along with a revised manual which incorporated those changes. For questions regarding this recall call 402-694-5003.
Quantity158
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K962288COVER-ALL, DRAPE-IT-ALL, CHAIR SLEVE, X-RAY SLEEVE, HEADREST COVER
Pinnacle Products, Inc., Jan 30, 1997