Recall Z-0955-2015 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare

The Revolution CT is a multi-slice (256 detector row) CT scanner consisting of a gantry, patient table, scanner desktop (operator console), system cabinet, power distribution unit (POU). and interconnecting cables.The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.

DateJan 26, 2015 (initiated Nov 12, 2014)
StatusTerminated
ReasonA required quality control test was not performed during installation associated with the software of the Revolution CT scanner.
Root causeRadiation Control for Health and Safety Act
ActionGE Healthcare Planned Action (s) to repair defect or to bring product into Compliance letter dated January 16, 2015 to GE Healthcare. 1. A GE Healthcare service representative will perform a noise specification and tolerance calibration to correct all affected systems via Field Action Instruction (FMI 25457). 2. A customer notification letter which includes a statement that GE Healthcare will without charge, remedy the defect or bring the product into compliance. CDRH approves the CAP subject to the following conditions: 1. Notification of all dealers and purchasers is to be made within 15
Quantity13 units
DistributionUS Distribution to the states of CA, FL, IL, NY, UT and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133705REVOLUTION CT
Ge Medical Systems, LLC, Apr 11, 2014