Recall Z-0987-2015 ยท JQP Calculator/Data Processing Module, For Clinical Use

Siemens Healthcare Diagnostics, Inc.

ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality

DateJan 15, 2015 (initiated Dec 16, 2014)
StatusTerminated
ReasonEnabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.
Root causeUnder Investigation by firm
ActionA Siemens service provider visited each site to verify and/or update customer configuration on the ADVIA Chemistry XPT and CentraLink Systems to ensure proper reporting of results. An Urgent Medical Device Correction letter was issued on December 16, 2014 to all affected customers. Customers were also provided a Customer Bulletin providing instructions on how configuration setting should remain.
Quantity7
DistributionDistributed in the state of WA.
FDA recordOpen on fda.gov