Recall Z-0966-2015 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare

GE Healthcare, Discovery IGS 730 & Discovery IGS 740. Discovery IGS 730 : The angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular & non vascular diagnostic and interventional procedures. Discovery IGS 740: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in gene

DateJan 14, 2015 (initiated Dec 10, 2014)
StatusTerminated
ReasonPotential unintentional system motion or rotation due to a positioning user interface issue with Discovery IGS 730 and 740 products.
Root causeSoftware design
ActionConsignees were sent a 12/10/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref# 12223 dated December 10, 2014. The letter was addressed to Hospital Administrators / Risk Managers, Managers of Radiology / Cardiology and Radiologists / Cardiologists. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information ( US call 800-437-1171, other countries contact your local GE Healthcare Service Representative).
Quantity24 ( 7 US, 17 OUS)
DistributionWorldwide Distribution-Distributed in the states of CO, CT, MS, NJ, NY, OH, and WA, and the countries of BELGIUM, CANADA, CHINA, FRANCE, GERMANY, JAPAN, MEXICO, PERU, UNITED ARAB EMIRATES, and UNITED KINGDOM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113403GE DISCOVERY IGS ANGIOGRAPHIC,FLUOROSCOPIC X-RAY SYSTEM
GE Medical Systems SCS, Feb 14, 2012
K133278DISCOVERY IGS 740
GE Medical Systems SCS, Apr 24, 2014