Recall Z-0983-2015 ยท OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

Biomet Spine, LLC

Timberline Anchored Lateral Fixed Retractable Sleeve Model No. 8630-0205. Timberline Anchored Lateral Fixed Retractable Sleeves are used with the Timberline MPF System to help establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae. The Timberline Retractable Drill and the Timberline Retractable Awl fit within this Sleeve. Spinal Surgical Device.

DateJan 15, 2015 (initiated Aug 22, 2013)
StatusTerminated
ReasonThe inner diameter of the Sleeve shaft is undersized, resulting in interference fit when inserting the Anchored Lateral Straight Retractable Awl or the Anchored Lateral Straight Retractable Drill into the Sleeve.
Root causeDevice Design
ActionThe firm, Lanx, sent a letter to their Consignees/customers dated October 15, 2013. The letter describes the product, problem and actions taken. Lanx and their distributors replaced the affected inventory with conforming product. Please contact Chief Compliance Officer at 303-501-8534 or by email: Randy.fesmire@lanx.com if you have any questions regarding this recall.
Quantity14 units
DistributionUS Distribution to states of: AZ, CA, NY, TN and TX including PR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123767LANX FUSION SYSTEM- SA
Lanx, Inc., Mar 18, 2013