Recall Z-1022-2015 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

INNOKAS MEDICAL OY

VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.

DateJan 28, 2015 (initiated Dec 3, 2014)
StatusTerminated
ReasonIf the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.
Root causeSoftware design
ActionA notification letter was sent to the distributors on 12/3/2014. The letter requests the devices be returned to the manufacturer for repair.
Quantity20 units
DistributionDistributed in the US and Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133810VC150 VITAL SIGNS MONITOR
Innokas Yhtyma, Oct 30, 2014