Recall Z-1015-2015 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

ASSY, POWER SUPPLY 130V 50HZ. This component is used with the following machines: 2008T Hemodialysis Machine, 2008K Hemodialysis Delivery Equipment, 2008 K2 Hemodialysis Delivery Equipment 2008K@home Hemodialysis Delivery System

DateJan 27, 2015 (initiated Dec 16, 2014)
StatusTerminated
ReasonUnqualified power assemblies shipped to customers.
Root causeUnder Investigation by firm
ActionConsignees were notified of the recall via letter sent on 12/18/2014 via certified mail. The letter informs consignees of the recall and requests that they take the following steps: Please check your stock for any 50 Hz Power Supplies (P/N 190092) with the name Electro Assemblies, LLC which is visible on the Transformer contained inside the Power Supply and discontinue use. Place the product in a secure, segregated area. Please contact your Fresenius Technical Services at 1-800-227-2572 option 4 then option 1, and reference the Field Action Number above. They will issue you a Return Good Aut
Quantity10
DistributionDistributed in OK and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121341FRESENIUS 2008T HEMODIALYSIS MACHINE WITH BIBAG SYSTEM
Fresenius Medical Care North America, Dec 6, 2012
K121421FRESENIUS 2008K HOME HEMODIALYSIS MACHINE WITH WIRELESS WETNESS DETECTOR (WETALERT)
Fresenius Medical Care North America, Design Cente, Jan 17, 2013
K994267FRESENIUS 2008K
Fresenius Medical Care, Mar 16, 2000