Recall Z-0972-2015 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis Q, and Artis Q.zen ceiling systems; model numbers: 10848281; 10848354: The Artis family is dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed include cardiac angiography, neuro angiography, general angiography, rotational angiography, mul

DateJan 14, 2015 (initiated Dec 3, 2014)
StatusTerminated
ReasonSuboptimal routing of the cable may result in increased wear over time. Without additional measures, damage of a cable may result in limited functionality and in rare cases, failure of a system. An ongoing procedure could be terminated in rare cases.
Root causeDevice Design
ActionSiemens sent an important customer safety information notice, dated December 3, 2014, to all affected customers. The letter identified the product, problem, and action to be taken. The following hardware modifications will be implemented with the corrective action AX042/14/S: 1. A modified cable outlet to provide additional space for the cabling. 2. A modified plastic cover at the cable outlet providing optimal cable guidance in the area of the arc. In general all systems will be checked for existing damage prior to the implementation of this action. For questions regarding this recall c
Quantity50
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123529ARTIS Q AND Q.ZEN- MODULAR ANGIOGRAPHIC SYSTEM
Siemens Medical Solutions USA, Inc., Feb 26, 2013