Recall Z-1120-2015 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems, Inc.

Philips Medical System Allura Xper X-Ray Angiographic

DateMar 2, 2015 (initiated Jun 6, 2014)
StatusTerminated
ReasonIn certain circumstances, a software error can lead to a situation where the five minute fluoroscopy audible signal does not sound.
Root causeRadiation Control for Health and Safety Act
ActionPhilips Medical Planned Action: 1. The affected customers will be sent a customer notification letter that identifies the software corrective actions and indicates that these corrective actions will be provided to the end-user free of charge 2. A mandatory Field Change Order with reference FCO72200270 will be released that requires Philips field service engineers to install Software release R7.2.8 which addresses the buzzer issue. The expected date of this FCO will be February 2015. The software release will contain updated timer function code to correct the intermittent fluoroscopy timer m
Quantity7439
DistributionWorldwide Distribution - USA nationwide including Puerto Rico, Austria, Azerbaijan, Australia, Algeria, Argentina, Albania, Brazil, Belgium, Bangladesh, Bahamas, Bulgaria, Cambodia, Chile, Canada, China, Croatia, Czech Republic, Netherlands, New Zealand, Norway, Oman, Pakistan, Palestine, Panama, P
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102005ALLURA XPER OR TABLE SERIES
Philips Medical Systems North America, Inc., Aug 9, 2010