Recall Z-0960-2015 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Symbia T and Symbia T2 computer tomography systems. These systems are used to perform SPECT, CT, and SPECT + CT scans and imaging studies.

DateJan 13, 2015 (initiated Mar 5, 2014)
StatusTerminated
ReasonThere are confirmed reports of a component coming loose inside the CT enclosure and damaging the system. components.
Root causeComponent change control
ActionThe firm, Siemens, sent an undated "Customer Safety Advisory Notice" letter to all customers. The described the product, problem and actions to be taken. The letters inlcuded instructions for customers to: 1) continue to use their systems while waiting for the field correction to be scheduled and performed; 2) ensure that the safety advisory is placed in the system's instruction for use; and, 3) closely observe the patient during studies. The letter also instructs customers of record to forward a copy of the letter to the new owner if the equipment was sold or transferred. Customers with an
Quantity295 systems
DistributionWorldwide Distribution: US Nationwide including Puerto Rico; Austria, Australia, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, Egypt, Finland, France, Germany, Great Britain, Hungary, Ireland, India, Iran, Italy, Japan, Mexico, Netherlands,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082506SYMBIA, VERSION 4.0
Siemens Medical Solutions USA, Inc., Nov 21, 2008