Recall Z-0219-2016 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Medical Systems, Inc.

HeartStart MRx monitor/defibrillator Models M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac

DateNov 5, 2015 (initiated Dec 23, 2014)
StatusTerminated
ReasonMRx monitor/defibrillator could reboot at an indeterminate time, potentially causing therapy to be interrupted or delayed
Root causeSoftware design
ActionPhilips Healthcare issued on 12/23/14 The Urgent Medical Device Correction notification/Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. A software upgrade will be provided free of charge for all units affected by one or more of these issues. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking customers to follow the Action to be T
Quantity75,693 units
DistributionWorldwide Distribution - US Nationwide and the countries of AFGHANISTAN, ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA AND HERZEGOWINA, BRAZIL, BRUNEI DARUSSALAM, BULGARIA, CAMBODIA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA (local name: H
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031187HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Aug 27, 2003
K040404HEARTSTART MRX, MODEL M3535A
Philips Medical Systems, Inc., May 4, 2004
K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A
Philips Medical Systems, Sep 8, 2005
K061707HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS
Philips Medical Systems, Aug 11, 2006
K062233HEARTSTART MRX MONITOR/DEFIBRILLATOR
Philips Medical Systems North America, Inc., Nov 22, 2006
K063375HEARTSTART MRX, MODELS M3535A OR M3536A
Philips Medical Systems, Jan 11, 2007