ClearanceBrief
Recall Z-1016-2015 ยท JAK System, X-Ray, Tomography, Computed

Mobius Imaging, LLC

Airo Mobile Intraoperative CT I Airo; Model #: MobiCT-32. Computed Tomography x-ray system.

DateJan 27, 2015 (initiated Dec 8, 2014)
StatusTerminated
ReasonThere is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve.
Root causeSoftware design
ActionMobius sent an Urgent Field Safety Notice dated December 8, 2014, to all affected customers. The letter explains the recall and provides advise on action to be taken by the user. The letter also informs consignees that Mobius Imaging is working on a software fix for the software bug/communication error. The final fix is anticipated in early 2015. Advise on action to be taken by the user: The user should be prepared for the potential necessity to perform a manual navigation registration in order to use navigation. In general, do not use navigation data transmitted by the AIRO Mobile CT Scann
Quantity21 devices distributed. 15 of these devices affected by the software v. 1.1.1+ patch
DistributionUSA including AK, IL, WA, MA, NY, WV, NC, TX, CA, PA and Internationally to Germany, Netherlands, Belgium, Switzerland, Saudi Arabia, Egypt.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131431AIRO CT SYSTEM
Mobius Imaging, LLC, Sep 26, 2013