Recall Z-0963-2015 ยท MKJ Automated External Defibrillators (Non-Wearable)
Philips Medical Systems, Inc.
HeartStart MRx Monitor/Defibrillator; Model. Numbers: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advan
| Date | Jan 13, 2015 (initiated Nov 19, 2014) |
| Status | Terminated |
| Reason | Device may malfunction, which could cause therapy to be delayed, disabled, or delivered inadvertently. |
| Root cause | Software design |
| Action | On November 19, 2014, Philips notified consignees of the recall via letter. The Urgent Medical Device Correction notification/Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. A software upgrade will be provided free of charge for all units affected by one or more of these issues. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking |
| Quantity | 74,499 devices |
| Distribution | Worldwide Distribution - US Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K031187 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Aug 27, 2003 |
| K040404 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Inc., May 4, 2004 |
| K051134 | HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A Philips Medical Systems, Sep 8, 2005 |
| K061707 | HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS Philips Medical Systems, Aug 11, 2006 |
| K062233 | HEARTSTART MRX MONITOR/DEFIBRILLATOR Philips Medical Systems North America, Inc., Nov 22, 2006 |
| K063375 | HEARTSTART MRX, MODELS M3535A OR M3536A Philips Medical Systems, Jan 11, 2007 |