ClearanceBrief
Recall Z-0971-2015 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Force; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

DateJan 14, 2015 (initiated Dec 3, 2014)
StatusTerminated
ReasonArtifacts are found in acquired imaging on the SOMATOM Force.
Root causeSoftware Manufacturing/Software Deployment
ActionA customer advisory notice, dated December 3, 2014, was sent to end users that identified the product, problem, and action to be taken. Customers were informed that new software would be installed via remote access.
Quantity7
DistributionPA, MA, MN, SC, MD, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133589SOMATOM FORCE
Siemens Medical Solutions USA, Inc., Apr 17, 2014