Recall Z-0586-2015 ยท MKJ Automated External Defibrillators (Non-Wearable)
Heart Sync, Inc.
ADULT Radiotranslucent Electrode, Part number T100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
| Date | Dec 24, 2014 (initiated Nov 11, 2014) |
| Status | Terminated |
| Reason | Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the |
| Root cause | Device Design |
| Action | The firm, HeartSync, sent an "Voluntary Field Safety Alert for Heart Sync Multi-function Defibrillator Electrodes" notice dated 11/11/2014 to customers via email notifying them of the incompatibility with the Philips FR3 and FRX AEDs and requesting response to the notice. On 12/1/2014 (dated 11/26/2014) Heart Sync issued a Voluntary Device Correction via Press Release. The Press Release included a description of the reason for the recall, affected product, and instruction that the Philips FR3 and FRx AED units should only be used with electrodes specified in the (Philips) equipment manuals. |
| Quantity | sold in boxes of ten electrodes: 4692.9 boxes; 46929 units total |
| Distribution | Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K080421 | HEART SYNC, MODELS C-100 AND T-100 Heart Sync, LLC, Feb 29, 2008 |