ClearanceBrief
Recall Z-0584-2015 ยท MKJ Automated External Defibrillators (Non-Wearable)

Heart Sync, Inc.

ADULT/CHILD Radiotransparent Electrode, Part number C100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.

DateDec 24, 2014 (initiated Nov 11, 2014)
StatusTerminated
ReasonHeart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
Root causeDevice Design
ActionThe firm, HeartSync, sent an "Voluntary Field Safety Alert for Heart Sync Multi-function Defibrillator Electrodes" notice dated 11/11/2014 to customers via email notifying them of the incompatibility with the Philips FR3 and FRX AEDs and requesting response to the notice. On 12/1/2014 (dated 11/26/2014) Heart Sync issued a Voluntary Device Correction via Press Release. The Press Release included a description of the reason for the recall, affected product, and instruction that the Philips FR3 and FRx AED units should only be used with electrodes specified in the (Philips) equipment manuals.
Quantitysold in boxes of ten electrodes: 116 boxes; 1160 units total
DistributionWorldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080421HEART SYNC, MODELS C-100 AND T-100
Heart Sync, LLC, Feb 29, 2008