Recall Z-1210-2015 ยท JAK System, X-Ray, Tomography, Computed

Xoran Technologies, LLC

Xoran MiniCAT X-ray imaging device

DateMar 2, 2015 (initiated Sep 8, 2014)
StatusTerminated
ReasonOut of tolerance kVp, Unexposed (black) frames, and Out of tolerance Dose.
Root causeRadiation Control for Health and Safety Act
ActionXoran Planned Action : A Xoran Technologies service representative will re-measure and perform calibration to correct all affected systems for kVp and dose issues. For the image quality issue, the service representative will mark the black frame and then attempt to reconstruct an adequate image. CDRH also understands that you are continuing your investigation to determine the root cause and a more permanent corrective action for the 3 identified defects. 1. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21
Quantity432
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113421MINICAT
Xoran Technologies, Inc., Jul 11, 2012