Recall Z-0984-2015 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
ARTISTE MV System, Linear Accelerator (LINAC) with SysVC10A software. The intended use of the ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
| Date | Mar 18, 2015 (initiated Nov 19, 2014) |
| Status | Terminated |
| Reason | There may be an existing dark current phenomenon on ARTISTE LINAC in combination with IMRT or mARC treatments using unflat beams. Software issue. |
| Root cause | Software design |
| Action | Siemens Planned Action: 1. Providing recommendations for user actions in terms of routine dosimetric quality control checks for any IMRT treatment at very high dose rates (Multiple X) as well as for mARC treatments. 2. Software update for the Function Controller #3, which controls parts of the beam generation. This software update enables the correct switching between Pulse Forming Network (PFN) and reduced Pulse Forming Network (iPFN) in pause state, to reduce the dark current during the movement as much as possible. This software will be installed by a Siemens service representative. 3. |
| Quantity | 3 |
| Distribution | US Distribution to the states of UT, WI, and NY. |
| FDA record | Open on fda.gov |