Recall Z-0739-2015 ยท KOG
Stryker Endoscopy
PKG, 5MM, 45CM, INSERT, FUNDUS GRASPER FORCEPS, P/N 0250080735 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used
| Date | Dec 12, 2014 (initiated Nov 17, 2014) |
| Status | Terminated |
| Reason | The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods |
| Root cause | Error in labeling |
| Action | Stryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Cu |
| Quantity | 180,573 units total |
| Distribution | Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.