Recall Z-0507-2015 ยท IYE Accelerator, Linear, Medical

TomoTherapy Incorporated

TomoTherapy Treatment System with software versions 2.0.1 / 2.0.2 / 2.0.3/ (Hi:Art 5.0.1 / 5.0.2 / 5.0.3)

DateDec 12, 2014 (initiated Nov 21, 2014)
StatusTerminated
ReasonAccuray has become aware of a potential safety issue related to the TomoTherapy Treatment System caused by a failure to monitor the jaw position after a jaw error occurs. This may result in an incorrect jaw position during treatment, without generating system interruption.
Root causeSoftware design
ActionAccuray sent an "Urgent Field Safety Notice" dated November 21, 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customers. For questions they can contact Accuray Customer Support at 1-877-668-8667 or visit the Accuray Technical Solutions Center at www.accuray.com>Services & Support or call: North America: +1.877.668.8667, Europe: +800.4141.9595, Hong Kong: +852.2247.8688, Japan: +81.3.6269.9556. Requested consignees to complete and return the Acknowledgement Form to Accuray Regulatory Affairs.
Quantity210 ( 72 USA; 138 OUS )
DistributionWorldwide Distribution - USA (nationwide) included the states of : AL, AZ, AR, CA, FL, GA, IL, IN, KS, ,Y, MD, MN, MO, NE, NM, NY, NC, OK, OR, PA, SD, TN, TX, UT, WA, WV, WI and WY., and the countries of : JAPAN, BELGIUM, NETHERLANDS, ITALY, GERMANY, KOREA, FRANCE, SWITZERLAND, ENGLAND, CANADA, MEXI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121934TOMOTHERAPY TREATMENT SYSTEM
Accuray Incorporated, Aug 29, 2012