Recall Z-0453-2015 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare, LLC

Innova 4100IQ Indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and Interventional procedures and optionally, rotational imaging procedure.

DateDec 16, 2014 (initiated Sep 4, 2014)
StatusTerminated
ReasonLoss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
Root causeDevice Design
ActionConsignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to t
Quantity6 (1 US, 5 OUS)
DistributionWorldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.