Recall Z-0452-2015 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare, LLC

Discovery IGS 740 Indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

DateDec 16, 2014 (initiated Sep 4, 2014)
StatusTerminated
ReasonLoss of imaging - affected devices may experience a loss of real-time interventional imaging and/or gantry motion due to the Electromagnetic (EM) sensitivity of the grid detection mechanism.
Root causeDevice Design
ActionConsignees were sent on 9/4/2014 a GE Healthcare "Urgent Medical Device Correction" letter GEHC Ref#12218 dated September 4, 2014. The letter was addressed to Director of Biomedical Engineering, Director of Radiology and Chief of Cardiology. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction, and Contact Information. Customers are to perform a system functional check as instructed in the letter. If the check fails, they should contact their local GE Healthcare Service Representative. GE Healthcare will correct all systems at no cost to t
Quantity1 - OUS
DistributionWorldwide Distribution -- US, including the states of CA, CO, FL, MO, PA; and, the countries of BELGIUM, CHINA, COLOMBIA, FRANCE, GREECE, MEXICO, and RUSSIA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133278DISCOVERY IGS 740
GE Medical Systems SCS, Apr 24, 2014