Recall Z-0835-2015 ยท GCJ Laparoscope, General & Plastic Surgery

Intuitive Surgical, Inc.

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Xi 4000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

DateDec 19, 2014 (initiated Nov 21, 2014)
StatusTerminated
ReasonThe storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.
Root causeLabeling design
ActionUrgent Medical Device Correction Letters were sent to customers on November 24, 2014.
Quantity1532
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141077DA VINCI FIREFLY IMAGING SYSTEM
Intuitive Surgical, Inc., Aug 12, 2014