Stryker Endoscopy

PKG, BABY METZENBAUM SCISSORS, CURVED, P/N 0250080028. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endosc

DateDec 12, 2014 (initiated Nov 17, 2014)
StatusTerminated
ReasonThe parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Root causeError in labeling
ActionStryker Endoscopy sent an Urgent Device Correction letter dated November 17, 2014, to affected customers. The letter informed the customers of the potential hazards and actions required. Customers were instructed to discard any old Instructions for Use for their Laparoscopic Manual Instruments and Accessories. Download the new Instuctions for Use (IFU1000-401-070 Revision J) from the URL link below: www.stryker.com/IFU1000401070 Customers were instructed to complete and sign the enclosed "Acknowledgement of Receipt" form and email them to endorecall@stryker.com or fax to (408) 855-6314. Cu
Quantity180,573 units total
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.