ClearanceBrief
Recall Z-0890-2015 ยท FPA Set, Administration, Intravascular

Roche Diagnostics Operations, Inc.

ACCU-CHEK Tender II 13/80 10 + 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a prop

DateDec 30, 2014 (initiated Nov 20, 2014)
StatusTerminated
ReasonRoche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Root causeDevice Design
ActionRoche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
QuantityTotal of 13,041 boxes
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972135MAERSK MEDICAL PURELINE COMFORT SUBCUTANEOUS INFUSION SET
P/L Biomedical, Aug 26, 1997