Recall Z-0831-2015 ยท GWM Device, Monitoring, Intracranial Pressure

Integra LifeSciences Corp.

Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for

DateDec 18, 2014 (initiated Nov 7, 2014)
StatusTerminated
ReasonDue to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
Root causeDevice Design
ActionIntegra sent Urgent- Voluntary Medical Device Recall Letters/Recall Acknowledgement and Return Form dated 11/7/2014 to customers and distributors via traceable courier service.
Quantity2,506 units
DistributionNationwide, Foreign
FDA recordOpen on fda.gov