Recall Z-0576-2015 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Syngo.plaza, Syngo.plaza VB10A, and syngo Imaging XS : Product Usage: Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images.

DateDec 15, 2014 (initiated Nov 11, 2014)
StatusTerminated
ReasonPrintouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of printer and the syngo.plaza printer configuration file. Printouts in anatomical size are not correct and may influence diagnostic decisions and/or therapy.
Root causeComponent design/selection
ActionSiemens sent a Safety Advisory notices letter dated November 11, 2014 to affected customers. The letter identified the affected product, problem, and actions to be taken. For questions call 1-888-826-9702.
Quantity149
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov