ClearanceBrief
Recall Z-0863-2015 ยท HQF Laser, Ophthalmic

Bausch & Lomb Inc

Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye.

DateDec 23, 2014 (initiated Nov 4, 2014)
StatusTerminated
ReasonA software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.
Root causeSoftware Manufacturing/Software Deployment
ActionBausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839.
Quantity79
DistributionWorldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133486STELLARIS PC VISION ENHANCEMENT SYSTEM
Bausch & Lomb, Mar 12, 2014