Recall Z-0473-2015 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

IntelliSpace Portal DX/HX/EX model number: 881001 IntelliSpace Portal DX/HX/EX Upgrade model number: 881011 IntelliSpace Portal DX/HX/EX Demo model number: 881012 IntelliSpace Portal IX model number; 881030 System, x-ray, tomography, computed

DateDec 1, 2014 (initiated Nov 9, 2014)
StatusTerminated
ReasonSoftware defect. In certain circumstances, the application may display incorrect measurements of cardiac and aortic anatomy used to select and size the Transcatheter heart valve (THV) to be implanted .
Root causeSoftware design
ActionConsignees were notified of the Field Safety Notice via certified mail.
Quantity32
DistributionWorldwide Distribution: US Distribution in states of: FL, NC, and VT; and the countries of Argentina, Austria, Belgium, Canada, Denmark, France, India, Italy, Japan, Netherlands, United Kingdom, and Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132524CT TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) PLANNING APPLICATION