Recall Z-0837-2015 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Mindray DS USA, Inc. dba Mindray North America

DPM 6 Patient Monitor and DPM 7 Patient Monitor. The product affected is the Mindray IBP module, P/N is 6800-30-50485 for use with Mindray DPM 6 and DPM 7 patient monitors . Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters.

DateDec 19, 2014 (initiated Oct 6, 2014)
StatusTerminated
ReasonMindray has identified an issue that involves the IBP module used with the DPM 6 and DPM 7 Monitor where the IBP measurement data may be displayed in an incorrect IBP tile on the monitors display.
Root causeProcess control
ActionMindray DS USA, Inc. sent an Urgent Field Corrective Action Letter dated 10/6/2014 to their customers via certified mail with return receipt. The letter identified the affected product, the problem, adverse effects on the patient, and corrective actions. Customers are asked to either verify that the IBP modules are not affected by following the steps provided in the "Procedures for Verifying IBP Channel Connections;" or to discontinue use of the IBP modules until a Mindray Service Representative contacts them and performs the test to verify the function of the IBP modules. Customers are to
Quantity455 units
DistributionDistributed to the states of AL, CA, CO, FL, GA, IA, IL, IN, LA. MA, MO, MS, MT, NC, ND, NY, OH, PA, SC, TX, VA, WA and WV.