Recall Z-0836-2015 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.

DateDec 19, 2014 (initiated Nov 7, 2014)
StatusTerminated
ReasonTwo issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Root causeSoftware design (manufacturing process)
ActionThe firm, IBA, sent a " User Notice" dated November 7, 2014, to the sole affected site-UPHS Customer. The notice describes the product, problem and actions to be taken. The customer was instructed that the KV/kV source acquisition mode has been disabled and shall not be used for patient alignment until system is officially confirmed to be back within tolerances. IBA will change the kV/kV Single Source acquisition workflow configuration file to be linked to the correct geometrical calibration file and will update the kV/kV Single Source calibration file that includes the colinearity parameters
Quantity1
DistributionUS Distribution in PA only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132847I2C
Ion Beam Applications S.A., Nov 20, 2013