Recall Z-0849-2015 ยท PFS System, Cystic Fibrosis Transmembrane Conductance Regulator Gene, Variant Gene Sequence Detection

Illumina Inc

Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific geno

DateDec 23, 2014 (initiated Nov 13, 2014)
StatusTerminated
ReasonIt was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.
Root causeSoftware design
ActionA customer letter dated November 13, 2014, was sent to all customers who is using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The letter informs the customers that Illumina has recently confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer. The customer notification outlines the potential issues, the results of investigation, actions required, and next steps. Customers are instructed
Quantity13 units
DistributionUS Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132750ILLUMINA MISEQDX CYSTIC FIBROSIS CLINICAL SEQUENCING ASSAY
Illumina, Inc., Nov 19, 2013