Recall Z-2094-2015 ยท NOU Continuous, Ventilator, Home Use

Breas Medical AB

Breas Vivo 50 Home Care Ventilator Product Usage: Respiratory Care

DateJul 16, 2015 (initiated Nov 7, 2014)
StatusTerminated
ReasonUnintended treatment termination could result from a keypad malfunction in some situations. The device erroneously interprets this as a Stop Treatment Instruction. An alarm will not sound, or be registered. Accessories and monitoring equipment connected to the Vivo 50 will stop functioning as the device enters stand-by mode.
Root causeDevice Design
ActionBreas Medical sent Urgent Release: Field Safety Notice letters dated November 7, 2014 and Field Safety Notice dated November 21, 2014 to their customers. The letters identified the affected product, problem and the actions to be taken.
Quantity846 US
DistributionWorldwide - United States Nationwide Distribution and the countries: Argentina, Belgium, Chile, Cyprus, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Malaysia, Netherlands, Norway, Panama, Poland, Portugal, Russia, Saudi Arabia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123144VIVO 50
Ge Healthcare, Breas Medical AB, Jun 18, 2013