Recall Z-0991-2015 ยท KHO Fluorometer, For Clinical Use

Tecan US, Inc.

Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The infinite 200 has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conforming to the AN

DateJan 15, 2015 (initiated Nov 10, 2014)
StatusTerminated
ReasonIncorrect lumi firmware version installed (E.027 instead of V2.00)
Root causeEmployee error
ActionTecan sent an Urgent Field Corrective Action letter dated October2014, to all affected consignees by via FedEx on/or about November 10, 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed not to use te kinetic luminescence measurement workflow requiring movement of th eplate transport out during kinectice measurements until the firmware has been upgraded to current version 2.0. A Tecan service representative will contact you to arrange a time to install the firmware. Additionally, they may want to review prior measurem
Quantity8 units
DistributionWorldwide Distribution - USA including MD and Internationally to Hong Kong, South Korea and the UK.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.