Recall Z-0204-2016 ยท MKJ Automated External Defibrillators (Non-Wearable)
Philips Medical Systems, Inc.
Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A
| Date | Nov 3, 2015 (initiated Nov 19, 2014) |
| Status | Terminated |
| Reason | 1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication |
| Root cause | Process control |
| Action | Philips Healthcare sent an " Urgent Field Safety Notice " dated November 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A hardware upgrade will be provided free of charge for all units affected by this issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: A Philips authoriz |
| Quantity | 1553 units |
| Distribution | Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Libyan Arab Jamahiriya, Lithuania, Mexico, Namibia, Netherlands, Palestine, Phil |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K031187 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Aug 27, 2003 |
| K040404 | HEARTSTART MRX, MODEL M3535A Philips Medical Systems, Inc., May 4, 2004 |
| K051134 | HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A Philips Medical Systems, Sep 8, 2005 |
| K061707 | HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS Philips Medical Systems, Aug 11, 2006 |
| K062233 | HEARTSTART MRX MONITOR/DEFIBRILLATOR Philips Medical Systems North America, Inc., Nov 22, 2006 |
| K063375 | HEARTSTART MRX, MODELS M3535A OR M3536A Philips Medical Systems, Jan 11, 2007 |