Recall Z-0820-2015 ยท LLZ System, Image Processing, Radiological

Carestream Health, Inc.

CARESTREAM DIRECTVIEW CR Software Generates digital mammographic images that can be used for screening and diagnosis of breast cancer.

DateDec 17, 2014 (initiated Oct 30, 2014)
StatusTerminated
ReasonReduced mammographic image quality when attempting to print true size multi-format images
Root causeSoftware design
ActionThe firm, Carestream Health Inc., sent an "URGENT: Medical Device Recall" notice dated October 28, 2014 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to use the one-up printing option rather the multi-format printing option if a hard copy is required. Carestream has a software update that will address this issue. A Carestream service representative will contact all customers to schedule a visit to correct this software issue. If you have any questions or concerns or would like to request the installation of the previous so
Quantity59 units
DistributionWorldwide Distribution: US distribution to states of : NY, OH, and NC; and countries of: Argentina, Brazil, China, Chile, Colombia, France, Greece, Hong Kong, India, Japan, Korea, Mexico, Peru, Spain, Taiwan, and Thailand.