Recall Z-0225-2015 ยท MAX Intervertebral Fusion Device With Bone Graft, Lumbar

Medacta Usa

The MectaLIF Oblique Handle, model number 03.22.10.0262, is used to insert the MectaLIF oblique spinal cage implants.

DateNov 18, 2014 (initiated Oct 21, 2014)
StatusTerminated
ReasonThe MectaLIF Oblique Handle has the laser marking "MEDIAL" on the incorrect side and could result in incorrect insertion of the device during surgery.
Root causeEmployee error
ActionMedacta sent an URGENT MEDICAL DEVICE RECALL letter dated October 21, 2014 were via e-mail to all sales representatives and distributors. The letters included instructions for the sales represetantives and distributors to: 1) check their inventory for the recalled devices; 2) immedaitely sequester the recalled devices and do not use them during surgery; 3) if the devices have been further distributed, inform them of the recall; and, 4) replacement devices will be provided. Customers with questions can contact Medacta USA at 805-910-6511.
Quantity32 oblique handles
DistributionWorldwide Distribution - US including CA, CO, ID and Internationally to Australia and Germany.