Recall Z-0523-2015 ยท MUE Full Field Digital, System, X-Ray, Mammographic

Siemens Medical Solutions USA, Inc

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

DateDec 6, 2014 (initiated Oct 28, 2014)
StatusTerminated
ReasonIt was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.
Root causeUse error
ActionA safety advisory notice, dated October 28, 2014, was sent to users that identified the product, problem, and actions to be taken to prevent risk to patients and users. Siemens technicians werre instructed to contact customers to schedule a field modification.
Quantity3
DistributionUS Nationwide Distribution in the state of: WV, NY, and FL..